1. Place a Integron® IT12 Chemical Integrator along with the material to be sterilized in a suitable package according to the recommended sterilization techniques. Place the indicators in those areas of most difficult access to the sterilizing agent.
2. Perform the sterilization cycle.
3. After the sterilization process has finished and evacuation/ventilation process is over, you may:
A. Remove the integrator indicator taking the necessary precaution to avoid contact with the remaining sterilizing agent.
B. Ventilate the load together with the integrator indicator and then perform the reading.
4. Analyze the results. If the SAFE integration condition has been reached, the sterilization process has been effective.